The Entube-Z generates a ±2V or ±1V scaled down version of its input signal, which can then be processed by most computer based measurement platforms.
0.1A | 0.5A | 1A | 2A | 3A | 4A | 5A | |
---|---|---|---|---|---|---|---|
Scaling Factor (1V output voltage) | 1:1 | 5:1 | 10:1 | 20:1 | 30: | 40:1 | 50:1 |
Overload Current (60Hz sinewave for 10s) | 0.5Arms | 2.5Arms | 5Arms | 10Arms | 15Arms | 20Arms | 25Arms |
Input impedance at 60Hz | 1 Ω | 200 mΩ | 100 mΩ | 50 mΩ | 333 mΩ | 25 mΩ | 20 mΩ |
Output impedance | 2k Ω |
Accuracy | ±0.2% (±0.4%) |
Bandwidth (-3dB point) | 40MHz |
Input-Output non-linearity | < 40 ppm/V |
Output voltage | ±100mV |
Gain temperature drift | ±25 ppm/°C |
Bandwidth (-3dB point) | BNC (Plug) |
Max total phase shift at 60Hz | < 0.05° |
Output Offset Voltage | < ±10μV |
Output connector | BNC (Plug) |
Encore is part of the family of current to voltage transducers designed to measure isolated and ground currents. The main focus of the Encore family is to ease and accelerate measuring the output current from low voltage current sources as well as isolated current transformers (CT).
Particular attention was paid on providing a solution that minimizes cost and improves overall performance on distributed systems such as substations, building monitoring, power distribution systems, etc. The Encore series covers the ranges of ±0.1A, ±1A, ±2A, ±3A, ±4A and ±5A with customizable bandwidth of up to 40kHz and 0.2% accuracy
Operating temperature | – 35 to 70 °C |
Storage temperature | – 40 to 80 °C |
Encore 5c operates as a current to voltage shunt that generates a ±100mV range signal for it’s nominal input current range. This signal can then be processed by a computer based measurement platform. In conjunction with the EasyConnect family, the Encore is the ideal solution for high channel density applications.
Due to its compact size and shape, the Encore can be easily mounted anywhere between the signal source and the data acquisition system. The versions up to 5A may even be used inline with the cable and not require any mounting at all. All can be secured to fixtures using cable ties.